4.3. TABULATED SUMMARY OF MEASURED VALUES

blue line

This section contains a summary of the measured values of dielectric properties of biological substances as a function of frequency and temperature, as reported in primary sources in the literature. Further information can be found in the listed references.
Tables 4.7 and 4.8 give the dielectric constant and conductivity values of different animal tissues, at indicated temperatures and frequencies. The temperature coefficient of the dielectric constant and conductivity of various body tissues are shown in Tables 4.9 and 4.10

Table 4.7
Real Part of the Complex-Dielectric Constant (Relative Permittivitiy) of Various Body Tissues
(Letters designate footnotes found at end of table.)

Table 4.8.
Conductivity (S/m) of Various Body Tissues
(Letters designate footnotes found at end of table.)

Table 4.9
Temperature Coefficient Of Dielectric Constant Of Various Body Tissues equation (Schwan, 1954)

Table 4.10.
Temperature Coefficient Of Conductivity Of Various Body Tissues equation (Schwan, 1954)

The complex permittivity is generally frequency dependent and at a frequency equation, can be described by the Debye equation, similar to Equation 4.1 (Schwan and Foster, 1980; Foster et al., 1980):
equation - (Equation 4.24)

where

equation
equation
equation
delta i = relative-permittivity change due to dispersion associated with epsilon i From Equation 4.24, we get

equation - (Equation 4.25)

and

- equation (Equation 4.26)

Hurt (1985) has written a Fortran program that performs a least-squares fit of Equations 4.25 and 4.26 to permittivity data. Figures 4.12 and 4.13 are the curves for relative dielectric permittivity and conductivity, respectively, that result from fitting a 5-term Debye relation to muscle data (Tables 4.7 and 4.8). For muscle, symbol is set equal to 4.3 (Grant et al., 1978; Hill, 1963).

The complex dielectric data in Tables 4.7-4.10 are from in vitro measurements of permittivity on excised human or animal tissues. Interest in reexamining tissue-permittivity values based on in vivo measurements has been increasing (Toler and Seals, 1977). With the in vivo procedure, a short monopole antenna is inserted into living tissue, and changes in the terminal impedance of the antenna are measured (Burdette et al., 1980). These impedance values can then be related to the complex permittivity of tissue.

Figure 4.12.
Relative dielectric permittivity for muscle (W. D. Hurt, private communication, USAF School of Aerospace Medicine, 1985).

Figure 4.13.
Conductivity for muscle (W. D. Hurt, private communication, USAF School of Aerospace Medicine, 1985).

Figures 4.14-4.16 compare in vivo data for dielectric constants and conductivity with in vitro data; Figures 4.17-4.19 give additional in vivo data. These figures show only slight differences between the in vivo and in vitro values, except in the case of canine fat tissue (Figure 4.17) where the in vivo dielectric constant values are a factor of approximately 1.5-2 times the reported in vitro results above 100 MHz. These differences in the dielectric constant are attributed primarily to possible differences in water content between the in vivo and in vitro measurement conditions (Burdette et al., 1980). Conductivity values are also generally higher than in vitro values found in the literature (Schwan, 1957).

Figure 4.14.
Measured values of relative dielectric constant of in vivo rat muscle and canine muscle (Burdette et al., 1980) compared to reference data (Schwan, 1975).

Figure 4.15.
Measured values of conductivity of in vivo rat muscle and canine muscle (Burdette et al., 1980) compared to reference data (Schwan, 1957).

Figure 4.16.
Measured values of relative dielectric constant and conductivity of in vivo and in vitro canine kidney cortex (Burdette et al., 1980) compared to reference data (Schwan, 1957).

Figure 4.17.
Measured values of relative dielectric constant and conductivity of in vivo canine fat tissue at 37ºC (Burdette et al., 1980).

Figure 4.18.
Measured values of relative dielectric constant and conductivity of in vivo rat brain at 32°C. Maximum SEM for epsilon' is indicated by error bars and SEM for epsilon" = ± 0.9 (Burdette et al., 1980).

Figure 4.19.
Measured values of relative dielectric constant and conductivity of rat blood at 23°C (Burdette et al., 1980).

Figures 4.20-4.24 show recently obtained in vivo data for several feline tissues at frequencies between 10 MHz and 1 GHz (Stuchly et al., 1981). These figures show significant differences in the properties of different types of the same tissue (e.g., skeletal and smooth muscle).

Figure 4.20.
Relative permittivity of cat smooth muscle in vivo: inverted delta and O show results obtained for two locations in the tissue. The vertical bars show the uncertainty due to the estimated measurement errors (SD) (Stuchly et al., 1981).

Figure 4.21.
Relative permittivity of cat spleen in vivo: inverted delta and O show results obtained for two locations; the vertical bars show the uncertainty due to estimated measurement errors (SD) (Stuchly et al., 1981).

Figure 4.22.
Average relative permittivity of two types of cat muscle in vivo (five samples for each point) (Stuchly et al., 1981).

Figure 4.23.
Average relative permittivity of cat internal organs in vivo (three to five samples for each point) (Stuchly et al., 1981).

Figure 4.24.
Relative permittivity of cat brain: tissue 1 -- gray matter; tissue 2 -- gray matter, 3 mm thick over white matter; tissue 3 -- white matter (Stuchly et al., 1981).

At frequencies below 100 kHz, a strong conductance anisotropy exists in muscle tissue (Rush et al., 1963). Data for anisotropic permittivity at low frequencies can be found in the works of Rush et al. (1963), Schwan (1957), and Johnson et al. (1975). Figure 4.25 shows the real part of the dielectric constant and conductivity of muscle tissue as a function of frequency for the parallel and perpendicular orientations (Epstein and Foster, 1983).

Figure 4.25.
The real part of the dielectric constant (a) and the conductivity (b) of the canine skeletal muscle tissue at 37°C as a function of frequency, in parallel orientation (open and closed circles) and perpendicular orientation (crosses and triangles), averaged over five measurements on different samples, The dotted lines are the data for non-oriented muscle tissue (Epstein and Foster , 1983).

Figures 4.26-4.28 show the real and imaginary parts of the dielectric constant and the conductivity of ocular tissue at 37°C as a function of frequency (Gabriel et al., 1983). Figures 4.29 and 4.30 compare the dielectric constant and conductivity of the normal and tumor mouse tissue as a function of frequency (Rogers et al., 1983).

Figure 4.26.
The real part of the dielectric constant, epsilon', of ocular tissues at 37°C (Gabriel et al., 1983). Key: solid block retina, empty circle iris, solid triangle choroid, empty square cornea, delta lens cortex, and solid circle lens nucleus.

Figure 4.27.
The imaginary part of the dielectric constant, epsilon", of ocular tissues at 37·C (Gabriel et al., 1983). Key: solid square retina, empty circle iris, solid triangle choroid, empty square cornea, delta lens cortex, and solid circle lens nucleus.

Figure 4.28.
The conductivity of ocular tissues at 37ºC (Gabriel et al., 1983). Key: solid square retina, empty circle iris, solid triangle choroid, empty square cornea, delta lens cortex, and solid circle lens nucleus.

Figure 4.29.
Real part of the dielectric constant, epsilon' , of normal and tumor mouse tissue as a function of frequency (Rogers et al., 1983). solid circle mouse muscle, solid square KHT fibrosarcoma tumor, solid triangle RIF/1 fibrosarcoma tumor.

Figure 4.30.
Conductivity of normal and tumor mouse tissue as a function of frequency (Rogers et al., 1983). solid circle mouse muscle, solid square KHT fibrosarcoma tumor, solid triangle RIF/I fibrosarcoma tumor.


blue line

Go to Chapter 5.1

Return to Table of Contents.

Last modified: June 24, 1997
© October 1986, USAF School of Aerospace Medicine, Aerospace Medical Division (AFSC), Brooks Air Force Base, TX 78235-5301

This is a Department of Defense computer system for authorized use only. DoD computer systems may be monitored for all lawful purposes, including to ensure that their use is authorized, for management of the system, to facilitate against unauthorized access, and to verify security procedures, survivability, and operational security. Using this system constitutes consent to monitoring. All information, including personal information, placed on or sent over this system may be obtained during monitoring. Unauthorized use could result in criminal prosecution.

POC: AFRL/HEDM, (210)536-6816,  DSN 240-6816 >