Table 7.7. Electrical Properties Of Simulated Muscle Tissue
For 13.56 - 2450 MHz At Three Temperatures (Chou et al., 1984)



15 ± 1°C
22 ± 1°C
30 ± 1°C
Frequency
(MHz)
Dielectric
Constant epsilon
'
Conductivity
sigma(S/m)
Dielectric
Constant epsilon '
Conductivity
sigma (S/m
)
Dielectric
Constant epsilon '
Conductivity
sigma (S/m)
2450 49.3 ± 1.3 2.09 ± 0.02 47.0 ± 0.9 2.17 ± 0.08 46.0 ± 0.3 2.43 ± 0.02
915 53.1 ± 0.7 1.10 ± 0.02 51.1 ± 0.6 1.27 ± 0.02 46.0 ± 0.3 2.43 ± 0.02
750 53.6 ± 0.6 1.09 ± 0.04 52.5 ± 0.6 1.26 ± 0.04 50.7 ± 1.0 1.43 ± 0.03
433 54.9 ± 0.7 1.07 ± 0.03 53.5 ± 0.5 1.21 ± 0.01 51.3 ± 0.8 1.38 ± 0.03
300 56.9 ± 0.6 1.06 ± 0.01 54.8 ± 0.7 1.17 ± 0.01 52.4 ± 0.7 1.36 ± 0.04
200 59.8 ± 0.6 0.91 ± 0.02 56.7 ± 0.7 1.06 ± 0.02 54.8 ± 1.0 1.21 ± 0.04
100 80.2 ± 1.2 0.76 ± 0.01 71.5 ± 1.1 0.89 ± 0.01 67.3 ± 1.3 1.03 ± 0.02
70 88.1 ± 0.7 0.67 ± 0.01 84.7 ± 0.5 0.76 ± 0.01 80.7 ± 2.8 0.92 ± 0.02
40.68 106.0 ± 2.0 0.60 ± 0.04 97.9 ± 3.8 0.70 ± 0.02 88.6 ± 4.7 0.84 ± 0.04
27.12 118.0 ± 5.0 0.51 ± 0.01 113.0 ± 3.0 0.62 ± 0.02 109.0 ± 1.0 0.72 ± 0.01
13.56 167.0 ± 4.0 0.52 ± 0.02 149.0 ± 3.0 0.62 ± 0.03 141.0 ± 2.0 0.77 ± 0.02

This is a Department of Defense computer system for authorized use only. DoD computer systems may be monitored for all lawful purposes, including to ensure that their use is authorized, for management of the system, to facilitate against unauthorized access, and to verify security procedures, survivability, and operational security. Using this system constitutes consent to monitoring. All information, including personal information, placed on or sent over this system may be obtained during monitoring. Unauthorized use could result in criminal prosecution.