U.S. Patent 8,378,879 (invented by Douglas L. Lewis et al., assigned to Johns Hopkins University) describes a Continuous Wave Phase Difference Locked Loop (CW-PDLL) designed for the non-contact, remote detection of physiological movement (such as chest displacement from heartbeat and respiration).


Operating Principle

  1. RF Front-End and Phase Difference Detection:  A transmitter (TX) driven by a Voltage Controlled Oscillator (VCO) emits a continuous wave (typically microwave/K-band, around 24 GHz)  The reflected signal is captured by two receiver channels (RX1 & RX2), commonly implemented via a dual-mixer / quadrature (I/Q) Gunnplexer.

  2. Phase Difference Locked Loop (PDLL):  The intermediate frequency (IF) signals pass through low-pass filters F1 and F2 into an instrumentation amplifier (IA1).  IA1 produces an Error Signal representing the phase difference.  This error is filtered by F4 and amplified by OA3 before being fed back to modulate the VCO frequency.  The loop locks by shifting the transmit wavelength to maintain a constant electrical phase relation between the transmitter, target, and receivers (acting like an "accordion").  The resulting VCO tuning voltage is proportional to the physical target displacement.

  3. Thermal Drift Integrator:  An active Miller integrator (OA2) senses the error signal and drives the reference pin (REF) of IA1, providing an ultra-low-frequency DC servo that tracks out slow temperature drifts and environmental changes without suppressing vital sign frequencies.

  4. 1/R2 Automatic Gain Control (Optional AGC System):  Because radar return power falls off with target range (1/R2 for retroreflective/large surfaces or 1/R4 for point scatterers), a 10 kHz pilot tone is superimposed onto the VCO tuning port.  The received 10 kHz tone amplitude is isolated by high-pass filter F3, amplified by OA1, converted to DC via an RMS-to-DC converter, and compared against an offset setpoint by OA4 to adjust the Variable Gain Amplifier (UGA).

  5. Output Filtering:  F5 extracts the composite physiological band.  F6 extracts the low-frequency respiration waveform, while F7 isolates the higher-frequency cardiac pulse.

Component Value Calculations for Unmarked Components

To determine realistic component values, standard design parameters for a 24 GHz homodyne biomedical radar tracking humans at 1 to 5 meters range are applied:

Thermal Drift Integrator

OA2 is configured as an inverting Miller integrator:

                    C_int
                 +---||---+
                 |        |
   ERROR ----+---/\/\/\---+
             |    R_int   |
             |           |\
             |     0V ---|+ \
             |           |   >--- To REF (IA1)
             |       +---|- /
             |       |   |/ OA2
             v       v
         (R_err)   GND

Cutoff Frequency Target:  Must reject steady-state DC and slow thermal drift while passing the lowest respiration frequency (fresp,min ≈ 0.1 Hz).  The corner frequency fc,drift is chosen at:  fc,drift ≈ 0.01 Hz to 0.02 Hz

Time Constant:

τint = Rint * Cint = 1/2πfc = 1/2π * 0.016 Hz ≈ 10 seconds

Calculated Values:

Main Loop Amplifier & Summing Node

OA3 is configured as a non-inverting amplifier following filter F4:

Av3 = 1 + (Rg3 / Rf3)
                      +---------------+
                      |               |
                      |    R_f3       |
               +------+-/\/\/\--+     |
               |                |     |
               |   |\           |     |
    F4 OUT ----+---|+ \         |     |
                   |   >--------+-----+---> To UGA & Summing Node
               +---|- / OA3
               |   |/
               \
               / R_g3
               \
               |
              GND

Gain Requirement:  Typical differential amplitude out of IA1 for millimeter displacements is ≈ 10 mV to 100 mV.  To drive the VCO varactor over a dynamic range of ≈ 0.5V to 2V, a loop gain Av3 ≈ 10 to 20 (20 dB to 26 dB) is required.

Calculated Values:

Summing Resistors at VCO Modulation Input:

The output of OA3 and the 10 kHz pilot oscillator are summed passively into the high-impedance varactor line of the VCO:

AGC Pilot Generator & Input Potentiometer

10 kHz oscillator level potentiometer (OSC LEVEL):

Pilot Recovery Stage

The received 10 kHz tone is tapped from RX2 and passed through high-pass filter F3 into non-inverting amplifier OA1:

                       R_f1
                 +---/\/\/\---+
                 |            |
                 |   |\       |
    F3 OUT ------+---|+ \     |
                     |   >----+---> To RMS-to-DC Converter
                 +---|- / OA1
                 |   |/
                 \
                 / R_fixed
                 \
                 |
                 [ GAIN CAL (Pot 42) ]
                 |
                GND

Gain Target:

Calculated Values:

Av1,min = 1 + 100 kΩ/51 kΩ ≈ 3
Av1,max = 1 + 100 kΩ/1 kΩ = 101 (≈ 40 dB)

AGC Error Amplifier & Filter Network (OA4)

Between the RMS-to-DC converter and OA4, a ripple filter smooths the extracted DC level:

τagc = Rrms * Crms  -->  Crms = 50 ms/10 kΩ = ≈ 4.7 µF

Offset Adjustment & Difference Amplifier (OA4)

Filter Specifications (F1 through F7)

The patent functional blocks specify exact roles for the active/passive filter blocks:

F1, F2 Low-Pass Fc ≈ 50 Hz Blocks out-of-band noise while rejecting the 10 kHz pilot tone. 2nd-order Sallen-Key: R1, R2 = 22 kΩ, C1 = 0.22 µF, C2 = 0.1 µF

F3 High-Pass Filter fc ≈ 8 kHz Passes only the 10 kHz pilot tone. Active 2nd-order high-pass: R1, R2 = 2.0 kΩ, C1, C2 = 10 nF

F4 Low-Pass Filter fc ≈ 50 Hz Identical to F1 and F2 (per patent specification). Eliminates high-frequency noise from entering the VCO drive.

F5 Band-Pass Filter 0.05 Hz−4.0 Hz Extracts composite vital sign band (breathing + cardiac). High-pass corner at 0.05 Hz 0.05 Hz (R = 1 MΩ, C = 3.3 µF); low-pass corner at 4.0 Hz (R = 39 kΩ, C = 1.0 µF)

F6 Low-Pass Filter fc ≈ 0.5 Hz Strips cardiac pulse, outputting clean respiration waveform. 2nd-order Butterworth: R1, R2 = 330 kΩ, C1 = 1.5 µF, C2 = 0.68 µF

F7 High-Pass Filter fc ≈ 0.8 Hz Rejects breathing baseline, isolating cardiac pulse (48-180 bpm). 2nd-order active HPF: C1, C2 = 1.0 µF, R1, R2 = 200 kΩ

Summary Checklist of Unmarked Passive Components in Figure 1

Drift Integrator

  Series Input Resistor to OA2: 1.0 MΩ
Feedback Integrating Capacitor: 10 µF
   Error Signal Shunt Resistor: 100 kΩ

VCO Driver (OA3)

      Input Resistor to F4: 10 kΩ
   Feedback Resistor (Rf3): 100 kΩ
     Ground Resistor (Rg3): 10 kΩ
   Summing Resistor to VCO: 1.0 kΩ

Oscillator Summing

   OSC LEVEL Potentiometer: 10 kΩ
   Summing Resistor to VCO: 47 kΩ

AGC Amplifier (OA1)

         Feedback Resistor: 100 kΩ
 Fixed Ground Leg Resistor: 1.0 kΩ
    GAIN CAL Potentiometer: 50 kΩ

AGC Detector & Error Amp (OA4)

   Post RMS-to-DC Resistor: 10 kΩ
Filter Capacitor to Ground: 4.7 µF
  OFFSET ADJ Potentiometer: 20 kΩ
Input Resistor from Offset: 100 kΩ
         Feedback Resistor: 100 kΩ
 Output Resistor & Pull-Up: 10 kΩ